Working Principle



Under the action of AC voltage, the dielectric consumes a portion of electrical energy, which is converted into heat energy and causes loss. This energy loss is called dielectric loss. When an AC voltage is applied to a dielectric, there is a phase angle difference Ψ between the voltage and current in the dielectric. The complementary angle δ of Ψ is called the dielectric loss angle, and the tangent of δ, tgδ, is called the dielectric loss tangent. The tgδ value is a parameter used to measure dielectric loss. The instrument's measurement circuit includes a standard circuit (Cn) and a test circuit (Cx). The standard circuit consists of a built-in high-stability standard capacitor and the measurement circuit, while the test circuit consists of the test object and the measurement circuit. The measurement circuit is composed of a sampling resistor, a preamplifier, and an A/D converter. Through the measurement circuit, the amplitudes and phases of the current in the standard circuit and the test circuit are respectively measured. Then, using a microcontroller with digital real-time sampling and vector operations, the capacitance value and dielectric loss tangent of the test object can be obtained.



The instrument has built-in anti-interference measures to ensure accurate measurement under external electric field interference.



Technical Parameters



1

Operating Conditions

-5℃∽40℃

RH<80%

2

Anti-interference Principle

Frequency Conversion Method

3

Power Supply

AC 220V±10%

Frequency Unrestricted

4

High Voltage Output

0.5KV∽10KV

Every 0.1kV

Accuracy

2%

   

Maximum Current

200mA

   

Capacity

1500VA

   

5

Self-Excited Power Supply

AC 0V∽50V/15A

45HZ/55HZ

6

Resolution

tgδ: 0.001%

Cx: 0.1pF

7

Accuracy

△tgδ:±(Reading*1.0%+0.040%)

△C x :±(Reading*1.0%+1.0PF)

   

8

Measurement Range

tgδ

Unrestricted

C x

15pF < Cx < 200nF

   
 

10KV

Cx < 60 nF

   
 

5KV

Cx < 100 nF

   
 

1KV

Cx < 200 nF

   

CVT Test

Cx < 200 nF

   

9

CVT Ratio Range

10∽10000

CVT Ratio Accuracy

0.1%

 

CVT Ratio Resolution

0.1

 

10

Dimensions

430(L)×330(W)×330(H)

11

Memory Size

80 groups, supports U-disk data storage

12

Weight

28 Kg