Instrument Parameter Characteristics



Instrument Operating Environment: 1) Temperature: 0-40℃ 2) Altitude: Use within 2Km above sea level 3) Relative Humidity: 40-80% Humidity Measurement Range:



1) Battery Voltage Measurement Range: 0-10V (RG-2002) Minimum Resolution 1mV 2) Current Measurement Range: 0-12A Minimum Resolution 1mA 3) Internal Resistance Measurement Range: 0-1000mΩ Minimum Resolution 0.1mΩ 4) Identification Resistance Measurement Range: 0.1-999.9K Minimum Resolution 0.1K 5) Capacity Measurement Range: 0-10000mAH Minimum Resolution 1mAH Test Speed: 1) Static Test (Test All Functions): 1-2 Seconds 2)



Capacity Test (IC Current Charge/Discharge): 3-4 Hours Measurement Accuracy: 1) Voltage Measurement Accuracy: ±(Result×0.1%+3mV) (Voltage 0-9.99V) 2) Current Measurement Accuracy: ±(Result×2.0%+3mA) (Current 0-1.99A) ±(Result×0.5%+3mA) (Current 2.00-12.00A) 3) Internal Resistance Measurement Accuracy: ±(Result×1%+1mΩ) 4) Identification Resistance Measurement Accuracy: 10KΩ±1% 5) Battery Capacity Measurement Accuracy: 10AH±2% Internal CNC Voltage Source Specifications: 1) Maximum Output Voltage: 10V(RG-2002) 2) Maximum Output Current: 2A 3) Ripple Voltage: <20Mv 4) Load Regulation: <10% 5) Response Time: 1S Internal CNC Electronic Load Specifications: 1) Maximum Voltage: 10V(RG-2002) 2) Maximum Current: 2A (Continuous) 10A (10 Seconds) 3) Maximum Power: 40W (Continuous) 80W (10 Seconds) Power Supply Voltage: 220V±10% 50Hz Power Consumption: Maximum 50W Instrument Dimensions: 300mm×300mm×100mm Instrument Weight: 4.0Kg