Fault Analysis



Fault Phenomenon

Cause Analysis

Troubleshooting Method

Remarks

No display upon power-on

1) Power not connected

Connect the power supply

When replacing the fuse, use the same model; do not substitute with other types.

2) The instrument's 5A fuse is not properly installed or is open

Reinstall the fuse or replace it

   

Output current cannot reach 200A

1) Test line too long, i.e., wire cross-section too small

Replace with wires of larger cross-section

 

No current output

1) Device under test is open circuit

Check the device

Eliminate the fault

 

2) Open circuit fault in the test loop

Check the test loop and eliminate the open circuit fault

   


Precautions



1. Before turning on the power switch, ensure the test button is released (non-test state) and the current adjustment knob is turned counterclockwise to the "zero" position.



2. The instrument should be placed indoors in a dry, well-ventilated area free from corrosive gases.



3. After testing, wait until the power is completely discharged before removing test leads and other equipment.



Transportation and Storage



1. Transportation



When the equipment needs to be transported, it is recommended to use the company's instrument packaging wooden case and shock-absorbing materials to avoid unnecessary damage during transit and prevent unnecessary losses.



When the equipment is transported without a wooden case, stacking is not allowed. When using the company's instrument packaging box, the maximum stacking layers allowed is two.



During transportation, the instrument panel should face upward.



2. Storage



The equipment should be stored indoors in a dry, dust-free, well-ventilated area free from corrosive gases. Without wooden case packaging, stacking is not allowed.



When storing, the panel should face upward. Place moisture-proof items under the equipment to prevent dampness.