FZY-3 Mining Stray Current Tester is a portable rectifier instrument with high sensitivity and multi-range protection circuit.Its characteristics are: wide measurement range, with a total of twenty ranges, capable of measuring AC/DC stray voltage and stray current separately, no external power supply needed, safe and reliable. It is suitable for testing the current and voltage generated by steel rails, water pipes, cables, etc. in underground coal mines, preventing premature detonation of electric detonators and other combustion and explosion accidents caused by stray current discharge, reducing the fire sources in the mine to the *lowest level.I. Purpose
FZY-3 Mining Stray Current Tester is a portable rectifier instrument with high sensitivity and multi-range protection circuit. Its characteristics are: wide measurement range, with a total of twenty ranges, capable of measuring AC/DC stray voltage and stray current separately, no external power supply needed, safe and reliable. It is suitable for testing the current and voltage generated by steel rails, water pipes, cables, etc. in underground coal mines, preventing premature detonation of electric detonators and other combustion and explosion accidents caused by stray current discharge, reducing the fire sources in the mine to the *lowest level.
I. Usage Methods and Precautions
1. Before measurement, screw the test rod wires into the two binding posts respectively, and adjust the meter pointer to the zero scale line.
2. Place the test rods connected to the binding posts on the two points to be measured (e.g., steel rail, water pipe, cable sheath, coal, earth, etc.). The two test points should be on different conductors, with a distance of about 2 meters. Normally, measure every 50-100 meters, or determine test points temporarily as needed. Set the range switch to the appropriate gear, and the meter reading, after conversion, gives the corresponding current value.
3. During measurement, ensure good contact between the hard alloy tip of the measuring probe and the object under test. The instrument should be placed horizontally to ensure accurate readings.
4. If no AC stray current is measured at a location, also test for DC stray current, and vice versa.
5. When measuring stray voltage values, the corresponding ranges are: 50mA corresponds to 0.2V, 250mA to 1V, 1A to 4V, 5A to 20V, 10A to 40V. (i.e., current value multiplied by 4 gives the voltage value)
6. When the magnitude of stray current is unknown, first set to the *maximum range, then gradually decrease until an appropriate measuring range is reached.
7. When the measured current exceeds 5A, the instrument should only be used for short-time measurement; when the measured current exceeds 10A, only for instantaneous measurement, to avoid burning out components and the meter.
8. When measuring DC stray current, if the meter pointer deflects in the reverse direction, swap the test leads.
9. When the instrument is not in use, set it to the "Off" position.
II. Product Completeness
1. Main unit: one set
2. Test leads: one pair
3. Product manual: one copy
4. Product certificate of conformity: one copy
Note: This instrument has a one-year warranty and lifetime maintenance.