Brief Introduction to Z-V Lightning Counter Calibration Instrument:
1. The battery has been changed to a rechargeable battery, avoiding the hassle of replacing batteries and the faults caused by it.
2. The voltage is divided into three adjustable levels to accommodate different models and brands of counters.
3. The packaging has been improved to an aluminum alloy case, which is sturdy and durable for consumer use.
I. Technical Parameters
1. Voltage: DC 900V, 1200V, 1600V three levels
2. Discharge current >100A/10US
II. Usage Instructions
1. Charge before use. To ensure effective charging and extend battery life, use the dedicated charger. It takes 3-4 hours to fully charge an empty battery. The red light on the charger indicates charging; the green light indicates charging is complete. Before use, connect the attached ground wire plug at the rear end, and clamp the ground wire clip to the earth.
2. According to the installation height of the counter, appropriately extend the telescopic discharge rod.
3. Press the red button to turn on the high voltage for about 1 second. When the indicator light flashes slightly, gently tap the connection end between the counter and the lightning arrester for testing.
4. After each tap, the discharge rod tip should be moved away from the counter. If repeated testing is needed, you do not need to release the button. After about 1-2 seconds, when the indicator light flashes again, you can tap again for testing.
5. Continuous testing may cause the tester to heat up; please allow appropriate intervals.
6. If the indicator light does not flash after pressing the button for more than 3 seconds, the battery needs to be charged.
7. Do not disassemble the tester arbitrarily. If the battery capacity significantly drops or charging efficiency is too low, it needs to be replaced. Please purchase the dedicated battery pack from our company.
Note: When the indicator light flashes slightly, no part of the body should touch the electrode parts, otherwise there is a risk of electric shock.