Heterodyne Dielectric Loss Tester Detailed Information



 



Dielectric Loss Tester Working Principle:



Under the action of AC voltage, the dielectric consumes part of the electrical energy, which is converted into heat energy causing loss. This energy loss is called dielectric loss. When AC voltage is applied to the dielectric, there is a phase angle difference Ψ between the voltage and current in the dielectric. The complementary angle δ of Ψ is called the dielectric loss angle, and the tangent of δ, tgδ, is called the dielectric loss tangent. The tgδ value is a parameter used to measure dielectric loss. The instrument's measurement circuit includes a standard circuit (Cn) and a test circuit (Cx). The standard circuit consists of a built-in high-stability standard capacitor and the measurement circuit, while the test circuit consists of the test object and the measurement circuit. The measurement circuit is composed of a sampling resistor, a preamplifier, and an A/D converter. By measuring the amplitude and phase of the current in the standard circuit and the test circuit separately, and then using a microcontroller with digital real-time acquisition and vector operations, the capacitance value and dielectric loss tangent of the test object can be obtained.

The instrument has built-in anti-interference measures to ensure accurate measurement under external electric field interference.



Heterodyne Dielectric Loss Tester Technical Parameters



1

Operating Conditions

-5℃∽40℃

RH<80%

2

Anti-interference Principle

Frequency conversion method

3

Power Supply

AC 220V±10%

Frequency unlimited

4

High Voltage Output

0.5KV∽10KV

Every 0.1kV

Accuracy

2%

   

Maximum Current

200mA

   

Capacity

1500VA

   

5

Self-excitation Power Supply

AC 0V∽50V/15A

45HZ/55HZ

6

Resolution

tgδ: 0.001%

Cx: 0.1pF

7

Accuracy

△tgδ:±(Reading*1.0%+0.040%)

△C x :±(Reading*1.0%+1.0PF)

   

8

Measurement Range

tgδ

Unlimited

C x

15pF < Cx < 200nF

   
 

10KV

Cx < 60 nF

   
 

5KV

Cx < 100 nF

   
 

1KV

Cx < 200 nF

   

CVT Test

Cx < 200 nF

   

9

CVT Ratio Range

10∽10000

CVT Ratio Accuracy

0.1%

 

CVT Ratio Resolution

0.1

 

10

Dimensions

430(L)×330(W)×330(H)

11

Memory Size

80 groups, supports USB data storage

12

Weight

28 Kg