The dielectric loss insulation test can effectively detect overall moisture, deterioration, and local defects in the insulation of electrical equipment. It is widely used in electrical manufacturing, installation of electrical equipment, handover, and preventive tests.



This instrument adopts variable frequency power supply technology, utilizing microcontrollers and electronic technology for automatic frequency conversion, analog-to-digital conversion, and data calculation, achieving strong anti-interference capability, fast test speed, high accuracy, and easy operation.



The variable frequency power supply uses a high-power switching power supply, outputting pure sine waves at 45Hz and 55Hz, with automatic voltage boosting, capable of providing up to 10kV voltage; it automatically filters out 50Hz interference, suitable for on-site testing in substations and other environments with high electromagnetic interference.



: Dielectric loss tester, anti-interference dielectric loss tester, fully automatic dielectric loss tester, variable frequency dielectric loss tester, variable frequency dielectric loss and capacitance tester, anti-interference dielectric loss and capacitance tester, fully automatic dielectric loss and capacitance tester



 



Product Parameters Product Parameters



Technical Parameters



Measurement Range Capacitance Value 3pF ~ 60000 pF/10kV, 60pF ~ 1μF/0.5kV
Dielectric Loss Range 0 ~ 100% unlimited, Resolution: 0.001%
Automatic identification of capacitance, inductance, and resistance test objects
Maximum Error Capacitance Accuracy ± (1.0%×reading ±1 pF)
Dielectric Loss Accuracy ± (1.0%×reading ±0.04%)
Resolution Capacitance Resolution Minimum resolvable 0.001 pF
Dielectric Loss Resolution Minimum resolvable 0.001 pF
Test Current Range 10μA~5A
Internal High Voltage Set Voltage Range 0.5~10kV
Maximum Output Current 200mA
Test Frequency 50Hz single frequency, 45/55Hz automatic dual frequency conversion
Power Supply 180V ~ 270VAC, 50Hz/60Hz ± 1% (mains or generator supply)
Working Environment Ambient Temperature 0 ~ 40℃
Ambient Humidity ≤ 90%RH, non-condensing
Dimensions 465×350×360mm 3
Weight 27.6kg



Note:

1. The product appearance is subject to the actual product. Technical parameters and models are subject to change without prior notice. Any unauthorized use or copying will be prosecuted.