2. Automatic program-controlled current source technology, with a total of 1000 current gears set in the current source, automatically controlled by the internal microcontroller according to the measured resistance, thereby achieving a relatively wide measurement range and optimal measurement state, without the need for manual switching of current gears.
3. Fast response speed, can directly switch the tap changer in the measurement state, the instrument will automatically prompt, and the new resistance value will be displayed soon.
4. Highly intelligent design, with ingenious and advanced function settings, can automatically judge faults such as loose connection or disconnection of test leads.
5. Complete protection functions, can reliably protect the instrument from the impact of back electromotive force, with automatic discharge indication function.
6. Can display measurement current and measurement time.
7. Intelligent power management technology can effectively reduce internal heating of the instrument.
8. Can store 120 measurement data, which will not be lost after power failure.
9. All Chinese menus and operation prompts, intuitive and convenient.
Product Features
1. Adopts brand-new power supply technology, with rapid measurement, compact size, and convenient use
2. Large-capacity CPU can store 255 groups of measurement results
3. Adopts typical four-wire measurement method to improve the accuracy of measuring resistance (especially low resistance)
4. Comes with a panel-mounted micro printer, which can print measurement results in Chinese
5. Stored results can be viewed and printed
Technical Specifications
1. Output current: 5A, 10A
2. Measurement range: 5A: 1mΩ~4Ω
10A: 1mΩ~1Ω
3. Resolution: 1μΩ
4. Test accuracy: 0.2 level
5. Operating temperature: -10℃~50℃
6. Operating humidity: 0~80%
7. Power supply: AC 220V±10% 50Hz±1Hz
Usage Parameters
>Measurement range: 0~2000Ω. (Six gears in total)
>Test current: 5A (20mΩ~2Ω); 0.5A (20Ω); 0.05A (200Ω); 0.005A (2000Ω).
>Resolution: 1μΩ (20mΩ gear); 10μΩ (200mΩ gear); 0.1mΩ (2Ω gear); 1mΩ (20Ω gear); 10mΩ (200Ω gear); 0.1Ω (2000Ω gear).
>Accuracy: 1mΩ~2000Ω (±0.5%)
>Total power consumption: ≤100W
>Operating temperature and humidity: 0℃~40℃; ≤85% RH
>Withstand voltage: AC1.5kV 50HZ 1min
>Power supply: AC220V 50 HZ (±10%)
>Dimensions: 350mm(L)× 260mm(W)× 140mm(D)
>Weight: approximately 3kg