Technical Parameters
Edit
1. Measurement range and accuracy of the instrument:
a. Capacitance measurement
(1) Capacitance measurement range: 0.1μF to 2,000μF;
(2) Capacity measurement range: 5 to 20,000 kvar;
(3) Measurement accuracy: ±1.0%;
(4) Resolution: 0 to 1.999μF ±1.0%rdg;
0 to 19.99μF ±1.0%rdg;
0 to 199.9μF ±1.0%rdg;
0 to 1999μF ±1.0%rdg;
b. Current measurement
(1) Current measurement range: 0 to 199.9mA;
0 to 1.999A;
0 to 19.99A;
0 to 199.9A
0 to 1000A;
(2) Measurement accuracy: ±1.0%;
2. Power supply:
a. Rated voltage: industrial frequency 220V±10%;
b. Rated frequency: 50Hz;
c. Rated output: 26V/500VA;
3. Normal working conditions of the instrument:
a. Ambient temperature: 0℃ to +40℃
b. Relative humidity: ≤90%
4. Display mode: large-screen LCD with full Chinese output,
TPμp-40 panel-type thermal printer
5. Dimensions/Weight: 320×360×160 mm / 8 kg
6. Working principle
The capacitance bridge adopts a bridge circuit structure, with the standard capacitor and the capacitor under test as two arms of the bridge circuit. When measuring the capacitance value of a capacitor, close the test switch K, and the test voltage is applied to both the standard capacitor and the capacitor under test simultaneously. The processor collects the current signals flowing through both via sensors and processes them to obtain the capacitance value of the capacitor under test.
Due to the synchronous sampling technology of the standard capacitor and the capacitor under test, it is not affected by power supply voltage fluctuations; moreover, the measurement process is fully automatic, avoiding errors caused by manual operation, thus featuring good stability, repeatability, and accuracy.
Error Values
Edit
| Capacitance measurement range: | Error value: |
| 0.020μF to 0.200μF | ±0.5% |
| 0.200μF to 2.000μF | ±0.5% |
| 2.000μF to 20.00μF | ±0.5% |
| 20.00μF to 200.0μF | ±0.5% |
| 200.0μF to 2000.μF | ±0.5% |
| Inductance measurement range | Error value |
| 0.10mH to 0.200H | ±0.5% |
| 2.000H to 20.00H | ±0.5% |
| 20.00H to 200.0H | ±0.5% |
| 200.0H to 2000.H | ±0.5% |
| Reactive power of inductor | 0 to 20.00Mvar | Error value | ±1% |
| Active power of inductor | 0 to 20.00kW | Error value | ±1% |
| Dissipation factor of inductor | 0 to 20.00% | Error value | ±1% |
| Resistance component of inductor | 0 to 10.00MΩ | Error value | ±1% |